Loading a CAN Data Definition File (Symbol File Load)
You can configure the CAN data extraction conditions by loading a CAN data definition file (an SBL file).
*
Data
that has been edited using Symbol Editor is assigned to the instrument’s sub channels 1 to 60 according to the
order in the definition list (you can change the order in the Symbol Editor’s definition list).
*
* An SBL file (.SBL extension) is a CANdb file (.dbc extension) that has been converted and edited into a
physical value/symbol definition file using YOKOGAWA’s free Symbol Editor software. You can obtain Symbol
Editor from the YOKOGAWA website (http://www.yokogawa.com/ymi/).
CANdb files (.dbc) are signal definition database files created using the CANdb or CANdb++ software
produced by Vector Informatik.
Configuring the Scales of All Sub Channels (All SubChannel
Auto Scale/Default Scale) - CAN
Configure the scales of all sub channels.
Auto Scale: The upper and lower display range limits are set automatically.
Default Scale: To the greatest extent possible within the settable range, the maximum and minimum values are
used to set the display range setting (which is explained below) of the specified sub channel.
Sub Channel Display Settings (Display) - CAN
Configure the display label, zoom magnification, scaling method, and display range settings of the specified sub
channel.
Waveform Vertical Position (Position)
Configure this setting when Value Type is set to Logic.
Zooming by Setting a Magnification (V Zoom)
Configure this setting when Value Type is set to Logic.
Selectable range: x0.1 to x3.33
Display Range (Upper/Lower)
Configure this setting when Value Type is not set to Logic.
Set the upper and lower limits of the waveform display range.
Selectable range: −30.000E+30 to +30.000E+30
Auto Scale (Auto Scale)
Configure this setting when Value Type is not set to Logic.
The upper and lower display range limits are set automatically.
Default Scale (Default Scale)
Configure this setting when Value Type is not set to Logic.
To the greatest extent possible within the settable range, the maximum and minimum values are used to set the
display range setting (explained earlier) of the specified sub channel.
Display Group (Display Group)
4 Vertical Axis