When (Shunt Cal) Exec is performed, the input value when the relay circuit is off is applied.
P1:Y
Set the value for when the relay circuit is off (normally 0).
P2:X
When the relay circuit is on and (Shunt Cal) Exec is performed, the input value when the relay circuit is on is
applied.
P2:Y
Set the strain value that corresponds to the shunt resistance when the relay circuit is on.
Strain
P2:Y
P2:X
Strain input
Shunt calibration
In the setup menu, set the strain value
corresponding to the shunt resistor to P2:Y.
Current measured
value*
Before execution
After execution.
The gain is corrected.
* Obtained automatically when shunt calibration is performed
Unit (Unit)
You can set the unit using up to four alphanumeric characters.
Executing Shunt Calibration (Shunt Cal Exec)
Executes shunt calibration.
Display Type (Display Type)
The same as the linear scaling settings for voltage measurement.
For details about shunt calibration, see appendix 4.
• When you execute shunt calibration, select an appropriate range so that the measured values will stay
within the range when the shunt-calibration relay circuit is on. The instrument attempts to perform shunt
calibration within the current range.
• An error message will appear if shunt calibration fails (because of out-of-range values or some other
reason). When this happens, change the range, and perform shunt calibration again.