4-17
IM DL350-01EN
• Linear scaling cannot be performed for measurements of temperature or acceleration.
• The following waveforms cannot be linearly scaled.
Accumulated waveforms (except for the most recent)
• You can configure linear scaling for each channel.
• The specified scaling coefficient A and offset B are retained even after linear scaling is set to OFF.
• Computations are performed on the linearly scaled values.
Copy (Copy to)
You can copy the various vertical axis and linear scaling settings from the current channel you are configuring to
other channels whose modules are of the same type as the source channel.
Destination Channel
Set the destination channel to a channel from CH1 to CH6.
Execute (Execute)
Select Execute to copy the settings.
Gain Adjustment (Gain Adjustment)
Gain adjustment is used to adjust measured voltages. The adjustment range is ±5%. The setting resolution is
down to the fourth decimal place on all modules that can use this function.
• The setting is retained even when the instrument is turned off.
• This function can be used on voltage measurement modules 720211 (HS100M12), 720250 (HS10M12),
and 720254 (4CH 1M16).
• The gain adjustment is reset to 1.0000 when Execute or ON is specified for DC offset adjustment.
• Gain adjustment is not possible when DC offset is set to ON.
4 Vertical Axis