4-51
IM DL350-01EN
One Shot Output Settings (One Shot Out Setup)
A single specified data frame or remote frame is output at the specified time.
Message Format
Select the message format.
STD: Standard format
XTD: Extended format
ID (Hex)
Set the message ID of the output data frame.
Standard format (11 bits): 0x000 to 0x7ff
Extended format (29 bits): 0x00000000 to 0x1fffffff
Frame
Select the frame type.
Remote: Remote frame
Data: Data frame
DLC
Set the byte size of the data area of the data frame.
Selectable range: 0 to 15
Setup
Set the above items and Data (Hex).
• Data (Hex)
Configure this setting when the frame type (Frame) is set to Data.
In hexadecimal notation, specify the data frame value to output.
One Shot Out
Output the data.
You cannot output the data when in the Port Setup menu, Listen Only is set to ON.
4 Vertical Axis