23-6
IM DL350-01EN
Others (Others)
Cursor Read Mode (Cursor Read Mode)
You can select whether to perform cursor measurements on P-P compressed display data or the data that has
been acquired in the acquisition memory.
• Display Data (Display)
Cursor measurements are performed on the display data.
• Acquisition (ACQ)
Cursor measurements are performed on sampled data in acquisition memory.
Setting the Destination That Data Is Saved to upon Action Execution (Action Folder Mode)
You can select the destination that data is saved to when actions are executed.
• ON: Data is saved to the folder that is automatically created with the date.
If the number of files in the save destination folder exceeds 1000, a new folder is automatically created with
the date and an incremented sequence number (000 to 999) as its name, and the data continues to be saved
in the new folder.
• OFF: Data is saved in the folder that you have specified.
• The maximum number of files that can be saved to a single folder is 1000. Make sure that there are no
files in the destination folder before you execute an action.
• If the measurement count (Acquisition Count) is set to a number greater than 1000, the measurement
cannot be started.
Backup at Power-Off (Backup Save Mode)
• ON: When an SD card is inserted, waveform data in memory is saved to the SD card when the power is
turned off.
• OFF: Waveform data in memory is not saved to the SD card when the power is turned off.
Beep on Error (Beep on Error)
• ON: A beep is generated when an error occurs.
• OFF: A beep is not generated even when an error occurs.
File Operation (File)
23 Other Features