9-4
IM DL350-01EN
Example of Logic Signal Measurement
When you measure logic waveforms using vertical cursors, the measured values of Y1 and Y2 are determined in
the manner shown below.
1
Cursor 1
Cursor 2
2
3
4
5
6
7
8
• When the cursor order is 1->8
Bit Y1: 01001010
Y2: 10110010
Hex Y1: 4A
Y2: B2
• When the cursor order is 8->1
Bit Y1: 01010010
Y2: 01001101
Hex Y1: 52
Y2: 4D
When No Bits Are Turned Off
Cursor 1
Cursor 2
1
2
4
5
7
8
• When the cursor order is 1->8
Bit Y1: 01-01-10
Y2: 10-10-10
Hex Y1: 16
Y2: 2A
• When the cursor order is 8->1
Bit Y1: 01-10-10
Y2: 01-01-01
Hex Y1: 1A
Y2: 15
When Some Bits Are Turned Off
You can set the display format (Bit or Hex), the cursor order (Cursor Order), and the bit order (Bit Order) using
Preference > Logic in the Utility menu.
9 Cursor Measurement