![YOKOGAWA ScopeCorder DL350 User Manual Download Page 198](http://html.mh-extra.com/html/yokogawa/scopecorder-dl350/scopecorder-dl350_user-manual_911346198.webp)
16-3
IM DL350-01EN
External Start (Remote)
You can perform GO/NO-GO determination and output the results in sync with an external signal applied to the
GO/NO-GO I/O terminal of the instrument.
• OFF: GO/NO-GO determination is not performed through the use of an external signal.
• ON: GO/NO-GO determination is performed through the use of an external signal.
Action (Action)
Editing a Waveform Zone (Edit Zone)
Number of the Waveform Zone to Edit
Select the number of the waveform zone you want to edit from the range indicated below. If a zone has already
been created for that number, the zone will be displayed. If no zone has been created for a number, select a
base waveform from the base waveform editing menu (New), and then edit the zone.
Zone 1 to Zone 6 (Zone1 to Zone6)
Editing the Source Waveform Zone (Zone Edit)
Editing a Base Waveform (New)
When you create a new waveform zone, you need to select the waveform that you will base it on (the base
waveform). Select a waveform whose display is on.
CH1 to CH4
1
, Math1 to Math2
1 You can select the channel of an installed module. On a 4-CH module, select sub channel 1 or 2. However,
you cannot select the channel of a logic module, 16-CH voltage input, 16-CH temperature/voltage input, CAN
bus monitor, CAN & LIN bus monitor, or SENT monitor module.
Specifying the Editing Range
Select the part of the base waveform that you want to edit.
• Whole (Whole): The whole waveform is within the editing range.
• Part (Part): A portion of the waveform is within the editing range.
Whole
Part
Editing range
boundary 1
(Time Range1)
Editing range
boundary 2
(Time Range2)
Zone Settings
When Edit is set to Whole, you can set the upper, lower, left, and right boundaries of the waveform zone. When
Edit is set to Part, you can set the upper and lower boundaries of the zone.
• Upper and Lower Limits (Upper/Lower)
Selectable range: ±10 div vertically from the base waveform
When Edit is set to Part, you can set the upper and lower boundaries of the area between Time Range1 and
Time Range2.
• Left and right sides (Left and Right): These settings can be configured only when Edit is set to Whole.
Selectable range: ±5 div from the center of the screen
• Time range 1 and time range 2 (Time Range1 and Time Range2): These settings can be configured only
when Edit is set to Part.
Selectable range: ±5 div on the time axis
16 GO/NO-GO Determination (Scope mode only)