• Start Bit
Specify the bit number of the data extraction starting point.
0 to 63
• Bit Count (Bit Cnt)
Set the number of data bits to extract. The range that you can specify varies depending on the data type.
Unsigned, Signed: 2 to 32
Float: The bit length that can be set is only 32.
Logic: 1 to 8
• Byte Order
Select the method (endian) to use to store the data in the acquisition memory.
MSB
63
56 55
8 7
0
msb
0
lsb
LSB
MSB: Most significant byte
LSB: Least significant byte
msb: Most significant bit
lsb: Least significant bit
• • • • •
• • • • •
• • • • •
Big Endian
7
LSB
7
0 15
55 63
56
MSB
MSB: Most significant byte
LSB: Least significant byte
msb: Most significant bit
lsb: Least significant bit
• • • • •
• • • • •
Little Endian
msb
0
lsb
• • • • •
7
• Value Type
Select the type of CAN data to extract.
Unsigned: Unsigned integer
Signed: Signed integer
Float: Real number
Logic: Boolean
You can monitor up to 60 sub channels with a single port.
If the input is turned on for all the sub channels and the bit length of each sub channel is 16 bits or less, you
can monitor all 60 sub channels.
However, the amount of memory is limited, so each time that the length of a sub channel is set longer than
16 bits, the instrument turns the input of sub channels off (sets them so that they cannot be monitored)
in decreasing order starting with sub channel number 60 until the memory usage is within the memory
limitations.
All Sub channels Factor/Offset(All SubChannel Factor/Offset)
Configure these settings when the data type is set to Unsigned, Signed or Float. The Factor and Offset values
that you set here are used to convert the extracted data to physical values.
Factor: Scaling coefficient (value per bit)
Offset: Offset value
Selectable range: −10.000E+30 to +10.000E+30
Unit
Using up to 16 characters, enter the unit to display for the waveform.
4 Vertical Axis