7.6 Turning the NULL Computation On and Off
Procedure
1.
Press
MEASURE
to display the following menu.
2.
Press the
Null
soft key to select On or Off.
Explanation
NULL Computation
When you turn NULL computation on, the NULL value will be set to the next measured
value. While NULL computation is on, the measured result is calculated by subtracting
the NULL value from the measured value.
<<Corresponding Command Mnemonic>>
:SENSe:NULL[:STATe] 1|0|ON|OFF
7-6