Changing Settings
While a program is executing, you can:
• Turn output off.
• Turn program repetition on and off.
• Turn measurement on and off.
• Turn NULL on and off.
• Turn storage on and off.
• Turn storage file creation on and off.
• Turn the display on and off.
Executing Programs
For details on executing programs, see section 2.4.
Operations When the Program Trigger Is Set to Norm
You can use the RUN, STEP, and HOLD keys and external triggers to control the
program execution. The program interval time specifies the execution interval for each
step in the program.
If you press the RUN or STEP key during program execution, a sampling error occurs.
Operations When the Program Trigger Is Set to M.End (Option)
When measurement completes, the GS200 proceeds to the next step in the program and
changes the source level. If you set the program trigger to measurement end (M.End),
and the measurement trigger to source change completion (Ready), the program
executes the source and measurement operations of each step in the minimum time
required.
The RUN, STEP, and HOLD keys and external triggers are disabled. Additionally, the
program interval setting is ignored.
Note
If you use a PC to copy or delete files on the GS200 during program execution, the timing of
the program operations may be affected.
<<Corresponding Command Mnemonic>>
:PROGram:RUN
:PROGram:STEP
:PROGram:HOLD
:PROGram:PAUSe
:PROGram:CONTinue
:PROGram:COUNt
<integer>
|MINimum|MAXimum
6.7 Executing Programs
6-13
IM GS210-01EN