
2.3 Source
Source Range
Source Range
The following figure indicates the range that the GS200 can generate.
200 mA
–32 V
32 V
–200 mA
Current
Voltage
Voltage Source Range (See section 5.2 for the procedure)
The following voltage source ranges are available.
V
I
30 V/200 mA
range
10 V/200 mA
range
1 V/200 mA
range
Range Source Range
Resolution Max. Load Current
10 mV
±12.0000 mV
100 nV
--------
*
100 mV ±120.000 mV
1 µV
--------
*
1 V
±1.20000 V
10 µV
±200 mA
10 V
±12.0000 V
100 µV
±200 mA
30 V
±32.000 V
1 mV
±200 mA
* In voltage source mode’s 10 mV and 100 mV ranges, because the GS200 uses a voltage
divider, the output resistance is approximately 2
Ω
. Therefore, these ranges are not suitable
when the GS200 is connected to a load that allows current to flow (a low-impedance
load). Depending on the load current, the output voltage may decrease. Also, four-terminal
connection does not function for these voltage ranges. You must use the two-terminal
connection instead. Connect a high-impedance load, a load that is sufficiently larger than the
output resistance.
2-6
IM GS210-01EN