Item
Specifications
Safety standards
1
Compliant standard
EN61010-1, EN61010-2-030
Measurement category Other (O) (250 Vpeak)
2
The overvoltage category (Installation category) CAT II
3
Pollution degree 2
4
Emission
1
Compliant standards
EN61326-1 class A
EN55011 class A, group 1
EN61000-3-2 class A
EN61000-3-3
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 Class A, Group 1
Korea Electromagnetic Conformity Standard (
한국
전자파적합성기준
)
This is a Class A product.
Operation of this product in a residential area may cause radio interference in which case the
user is required to correct the interference.
•
Test conditions
1 V range, 1 V source, limit: ±200 mA, load resistance: 10
Ω
, and local sense (2W)
All other settings at factory default.
Cable conditions
•
Output terminals
Use the included measurement lead set (758933; red and black, one pc. each; length: 1 m)
to connect the OUTPUT terminals.
We recommend that you use the same cables when connecting the SENSE terminals for
four-terminal connection. (Use cables that are 1 m in length or less.)
•
BNC IN/OUT terminals
Use cables that are at least equivalent to 1.5D-QEW cables when connecting the BNC
terminals.
5
•
ETHERNET port
Use shielded (STP) LAN cables that are category 5 or better when connecting the
ETHERNET port.
(Use cables that are 30 m in length or less.)
•
USB port
Use cables that support USB 2.0 (Hi-speed USB) when connecting the USB port
5
•
GP-IB connector
Use GP-IB cables (IEEE488) when connecting the GP-IB connector.
•
RJ-11 IN/OUT terminals
Use shielded cables (758960, length: 1 m) when connecting the RJ-11 terminals.
Immunity
1
Compliant standard
EN61326-1 Table2
6
•
Immunity influence
Within the range specified by measured value ± 20% of range
•
Test conditions
1 V range, 1 V source, limit: ±200 mA, load resistance: 10
Ω
, and local sense (2W)
All other settings at factory default (the cable conditions were tested under the same
conditions as the emission test).
1 Applies to products that have a CE mark on their rear panel. For information on other products, contact your nearest
YOKOGAWA dealer.
2 The measurement category of the GS200 signal input terminals is Other (O). Do not use it to measure the main power supply
or for Measurement Categories II, III, and IV.
Measurement category Other (O) applies to measurement of circuits that are not directly connected to a main power supply.
This category applies to measurement of secondary electric circuits in equipment across a transformer. The estimated
transient overvoltage that may be applied to the GS200 signal input terminals is 1500 V.
Measurement category II applies to measurement of circuits, such as household electric appliances and portable electric tools,
that are connected to low-voltage installations.
Measurement category III applies to measurement of facility circuits, such as distribution boards and circuit breakers.
Measurement category IV applies to measurement of power source circuits, such as entrance cables to buildings and cable
systems, for low-voltage installations.
3 The overvoltage category (installation category) is a value used to define the transient overvoltage condition and includes the
rated impulse withstand voltage. The overvoltage category II applies to electrical equipment that is powered through a fixed
installation, such as a switchboard.
4 Pollution Degree applies to the degree of adhesion of a solid, liquid, or gas which deteriorates withstand voltage or surface
resistivity. Pollution Degree 2 applies to normal indoor atmospheres (with only non-conductive pollution).
5 Use cables of length 3 m or less.
6 Immunity test requirements for equipment intended for use in industrial environments.