7-2
IM CA700-01EN
Adjusting the DC Voltage
2.
Use the ▲ or ▼ key to select
Voltage Source
, and press
ENTER
.
The DC voltage source adjustment screen appears. The CA700 generates 0 VDC.
DC voltage source value
DC voltage source value
measured inside the CA700
3.
Adjust the DC voltage at two points: 0 V and 5 V. Adjust the CA700 DC voltage source value so
that it matches the reading on the digital multimeter.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number. You can adjust the voltage in the range of –0.05000 V to 0.05000 V.
4.
Press
Done (F5)
. The 0 V source value is adjusted.
The 5 V adjustment screen appears.
DC voltage source value
DC voltage source value
measured inside the CA700
5.
Like step 3, adjust the 5 V source value. The adjustment range is 4.50000 V to 5.50000 V.
6.
Press
Done (F5)
. The 5V source value is adjusted.
The 0 V internal measurement adjustment screen appears.
DC voltage source value
DC voltage source value
measured inside the CA700
7.1 Adjusting the Source Function