5-6
IM CA700-01EN
[BSN command] Queries the serial number.
BSN
Queries the serial number.
Normal Adjustment
Command = BSN?<CRLF>
→
Answer = xxxxxxxxx<CRLF>
Parameters
xxxxxxxxx : serial number
Yes
Yes
[CD command] Sets the source value for adjustment.
CD
Sets the source value for adjustment.
Normal Adjustment
Command = CD<CRLF>
→
Answer = CD<CRLF>
In the source adjustment condition, the current output settings are set
as the source adjustment values for the selected function, range, and
scale (+FS/0).
When the function is set to measurement, ERR13 is returned.
No
Yes
[CL command] Queries the adjustment item.
CL
Queries the adjustment item.
Normal Adjustment
Command = CL?<CRLF>
→
Return = CLm<CRLF>
Parameters
m = 3: Source adjustment
4: Measurement adjustment
No
Yes
[CMF command] Sets or queries the measurement function for adjustment.
CMF
Sets or queries the measurement function for adjustment.
Normal Adjustment
Command = CMFm<CRLF>
→
Answer = CMFm<CRLF>
Command = CMF?<CRLF>
→
Return = CMFm<CRLF>
Parameters
m = 0: DCV
12: DCmA
13: LOOP POWER
14: Pressure
15: Switch Test
When the function is set to measurement, ERR13 is returned.
No
Yes
[CMR command] Sets or queries the measurement range for adjustment.
CMR
Sets or queries the measurement range for adjustment.
Normal Adjustment
Command = CMRm<CRLF>
→
Answer = CMRm<CRLF>
Command = CMR?<CRLF>
→
Return = CMRm<CRLF>
Parameters
[DCV]
1: 5 V
3: 50 V
[DCA]
m = 0: 20 mA
1: 100 mA
[LOOP POWER] m = 0: 20 mA
[SWITCH TEST] m = 0: 2 kΩ
When the function is set to measurement, ERR13 is returned.
No
Yes
5.4 Commands