7-5
IM CA700-01EN
Adjusting the CA700
1
2
3
4
5
6
7
Index
Adjusting the DC Voltage Measurement
2.
Use the ▲ or ▼ key to select
Voltage Measure
, and press
ENTER
.
The DC voltage measurement adjustment screen appears.
Input voltage
Switch the measurement range.
Adjusting the 5 V Range
3.
Check that Input 5V is displayed, and apply 5 VDC from the DC voltage/current generator.
4.
Press
Done (F5)
. The 5 V range is adjusted.
Adjusting the 50 V Range
5.
Press
Range (F1)
to set the measurement range to 50 V.
Input 50 V appears.
6.
Check that Input 50V is displayed, and apply 50 VDC from the DC voltage/current generator.
7.
Press
Done (F5)
. The 50V range is adjusted.
Finishing the Adjustment
8.
Press
ESC
. The screen shown in step 1 appears.
7.2 Adjusting the Measurement Function