3-20
IM CA700-01EN
Saving the Data after Adjustment
9.
In the calibration display, press
As Left (F5)
.
A screen for monitoring the pressure applied to the pressure switch appears.
10.
Using the same procedure as step 4 and 5, save the calibration data after adjustment.
Explanation
Calibration Data File
The pressure switch calibration data before and after adjustment is saved to CSV files.
File Name
Calibration data before adjustment: SWxxF_y.CSV
Calibration data after adjustment:
SWxxL_y.CSV
xx: 00 to 99 (calibration group number)
Same number between the calibration data file before and
after adjustment
y: 0 to 9, sequence number assigned to data files before
adjustment and after adjustment
Because the number xx is the same for calibration data files before and after adjustment, the
association is maintained.
If you change the device information or calibration procedure, the number xx is incremented.
Data That Is Saved
Calibration conditions, device information, date and time of calibration, the pressure that opened
the pressure switch, the pressure that closed the pressure switch, the difference between the two
pressures, and the resistance when the pressure switch closed are saved.
3.6 Calibrating a Pressure Switch