3.2 Select the Calibration Procedure
Procedure
1.
From the top menu, use the ▲ or ▼ key to select
Calibrate
, and press
ENTER
.
The calibration display appears.
Information about the selected calibration
procedure (source function, measurement
function, number of calibration points,
and calibration direction)
2.
Press
Select Proc.(F2)
.
A list of registered calibration procedures is displayed.
Up to 20 sets of procedures can be registered. They are displayed in four windows.
3.
Press or
to change the page and ▲ or ▼ to select the calibration procedure you want to
edit. Press
ENTER
.
Information of the selected calibration procedure is displayed.
Explanation
Calibration Procedure
You can register up to twenty sets of calibration processes. A calibration procedure consists of a set of
device information, measurement function and source function.
To calibrate a pressure device, select one of the registered procedures to perform calibration.