3-6
IM CA700-01EN
Setting Sensor Type and Range (for external pressure sensor)
The Sensor Type is fixed to 16 MPa.
The procedure for setting Range is the same as in pressure measurement. See section 2.1.
Setting Unit (for Pressure), Averaging, and Scaling
These items are the same as with pressure measurement. See section 2.1.
Setting the 0% Value and 100% Value
10.
Use the ▲ or ▼ key to select
0% Value
, and press
ENTER
.
11.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
Likewise, set the 100% value.
Setting the Tolerance for Pass/Fail Judgment
12.
Use the ▲ or ▼ key to select
Tolerance
, and press
ENTER
.
13.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
The values assigned to 0% and 100% will be used as references.
Setting Source Conditions
14.
Hold down the ▼ key.
The display switches to page 3/3 (or 3/4 when the source function is set to EXT P, source condition setting
display).
When the source function is set to mA SOURCE,
mA SIMULATE, or V MEASURE
When the source function is set to P
Turns averaging on and off
Turn scaling on and off and set scaling.
Set the source function.
Set the 0% value.
Set the 100% value.
Set the number of calibration points.
Confirms the settings
Confirms the settings
Set the unit.
Set whether to calibration from 0% or 100%.
Set the source function.
Set the 0% value.
Set the 100% value.
Set the number of calibration points.
Set whether to calibration from 0% or 100%.
3.3 Setting the Device Information and the Calibration Conditions of the Pressure Transmitter