Source and Measure
2-7
IM CA700-01EN
1
2
3
4
5
6
7
Index
Scaling
By assigning scale values to the pressure at the lower span limit and the pressure at the upper span
limit, you can view measured values converted in any physical quantity.
Measured value (span)
Upper limit (Upper)
Lower limit (Lower)
Upper limit (Upper)
Lower limit (Lower)
Scale value
When scaling is set to ON, “SCL” is displayed on the screen.
Alarm
When a measured value exceeds the specified range, an alarm mark will appear, and the value will
blink. If you change the scaling settings while the alarm is set to ON, the alarm is automatically turned
off, and the upper and lower alarm limits are initialized.
Percentage Display
By taking the specified span to be 100%, the CA700 displays current values as percentages.
Reference Value and Relative Display
The CA700 displays the difference relative to the specified value.
Measure Value: The measured value when you press RELATIVE will become the reference value.
Setting Value: The reference value is set to any value of your choosing in the range of –999999 and
999999.
You can turn relative display on and off using a panel key.
Initializing the Settings
All settings except Function are initialized.
Averaging, Scaling and Alarm are set to OFF; Averaging and %Span are set to ON; Relative is set to
Measure Value.
Saving Measured Data
Measured data, including the measure and source values of Display2, are saved to a CSV file.
The following two save methods are available.
Save:
Measured values are appended to a file each time you press the Save function key. Data
is saved to the same file until any of the following conditions is met.
• When a setting is changed
• When the measuring range is changed
• When the number of save data points exceeds 2000
• When the power is turned off
The file name is SAVExx.CSV (xx: a sequence number between 00 and 99).
Logging: If you press the Logging Start function key, the CA700 will save data at the specified
interval. The operation stops when you press the Logging Stop function key or when the
specified number of data values are saved.
The file name is LOGxx.CSV (xx: a sequence number between 00 and 99).
You can set the measurement data save method using Data Save Setting under Device Setting from
the top menu. For details, see section 4.1.