3-18
IM CA700-01EN
3.6 Calibrating a Pressure Switch
Procedure
Selecting the Calibration Procedure
1.
In the calibration display, press
Select Proc. (F2)
.
The calibration procedure setting display appears.
2.
Use the or
key to change the page, and the ▲ or ▼ key to select a pressure switch (P
SWITCH) calibration procedure,and press
ENTER
.
The calibration procedure is selected, and the calibration display returns.
Switches to the pressure
switch calibration procedure
information