7-7
IM CA700-01EN
Adjusting the CA700
1
2
3
4
5
6
7
Index
Adjusting the mA Loop Measurement
2.
After step 1, use the ▲ or ▼ key to select
Loop Power
, and press
ENTER
.
The mA loop measurement adjustment screen appears.
Input current
3.
Generate a sink current of 20 mA from the DC voltage/current generator. The CA700 will
generate 24 V.
4.
Press
Done (F5)
. The 20mA loop measurement is adjusted.
Finishing the Adjustment
5.
Press
ESC
. The screen shown in step 1 appears.
Adjusting the Pressure Switch Resistance Measurement
2.
After step 1, use the ▲ or ▼ key to select
Switch Test Measure
, and press
ENTER
.
The pressure switch resistance measurement adjustment screen appears.
Resistance
3.
Adjust 0 Ω. Set the reference resistance to 0 Ω.
4.
Press
Done (F5)
. The resistance measurement is adjusted.
5.
Adjust 2 kΩ. Set the reference resistance to 2 kΩ.
6.
Press
Done (F5)
. The resistance measurement is adjusted.
Finishing the Adjustment
5.
Press
ESC
. The screen shown in step 1 appears.
7.2 Adjusting the Measurement Function