7-8
IM CA700-01EN
Explanation
Adjustment Points
The adjustment points of the measurement function are provided below.
Range
Generator Output
Adjustable Range
Lower Limit
Upper Limit
Voltage Measure
5.000 V
5.0000 V
4.500 V
5.500 V
50.00 V
50.00 V
45.00 V
55.00 V
Current Measure
20.000 mA 20.000 mA
18.000 mA
22.000 mA
100.00 mA 100.00 mA
90.00 mA
110.00 mA
Loop Power
20.000 mA 20.000 mA
18.000 mA
22.000 mA
Switch Test Measure
0 kΩ
0 kΩ
–0.200 kΩ
0.200 kΩ
2 kΩ
2.000 kΩ
1.800 kΩ
2.200 kΩ
7.2 Adjusting the Measurement Function