4.1 Setting How to Save Measured Data
1
2
3
4
5
6
7
Index
Saving and Deleting Data
Explanation
Save Method
The save method specifies how to save measured pressure, DC current, and DC voltage data.
Save:
Measured values are appended to a file each time you press the Save function key. Data
is saved to the same file until any of the following conditions is met.
• When a setting is changed
• When the measuring range is changed
• When the number of save data points exceeds 2000
• When the power is turned off
The file name is SAVExx.CSV (xx: a sequence number between 00 and 99).
Logging: If you press the Logging Start function key, the CA700 will save data at the specified
interval. The operation stops when you press the Logging Stop function key or when the
specified number of data values are saved.
The file name is LOGxx.CSV (xx: a sequence number between 00 and 99).
Note
• If the save method is set to Logging and the battery power in the CA700 is low when you start data
logging, the message “Continue Logging?” will appear. To proceed, press ENTER. To cancel, press ESC.
• If the battery power drops so much that the CA700 cannot continue data logging, it will show the message
“System has Stopped Logging” and stop data logging. The data logged up to that point will be saved.
Before you start data logging, check that there is sufficient battery power.
Save Interval
This is the data save interval that is used when the save method is set to Logging.
You can select 1 s, 2 s, 5 s, 10 s, 30 s, or 60 s.
Number of Data Values to Save
This is the number of data values to save when the save method is set to Logging.
You can set the value between 1 and 2000.
When the CA700 saves the specified number of data values, it will stop saving.