Source and Measure
2.1 Measuring Pressure
2-5
IM CA700-01EN
1
2
3
4
5
6
7
Index
33.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
Likewise, set Span Upper.
Note
The decimal point position is the same as the span decimal point position in Scaling.
34.
After entering the values, press
Setting Done (F5)
.
To discard all of the % Span settings, press the
ESC
key.
The system returns to the Display1 Setting display.
Note
The settings are not yet confirmed. To confirm the settings, press
Setting Done (F5)
on the Display1 Setting
display.
Setting the Reference Value and Relative Display
35.
Press the ▲ or ▼ key to select
Relative
.
Measure Value and Setting Value appear on the function menu.
36.
Press
Measure Value (F1)
or
Setting Value (F2)
.
If you select Measure Value, the measured value when you press RELATIVE will become the reference
value.
37.
If you select Setting Value, press
ENTER
.
The reference value setting screen appears.
Set the lower span limit.
Set the upper span limit.
Confirms the settings
38.
Press the or
key to move to the digit you want to change, press the ▲ or ▼ key to set the
number.
To confirm the value, press
ENTER
.
39.
After entering the values, press
Setting Done (F5)
.
To discard all of the relative settings, press the
ESC
key.
The system returns to the Display1 Setting display.
Note
The settings are not yet confirmed. To confirm the settings, press
Setting Done (F5)
on the Display1 Setting
display.
Initializing the Settings
37.
Use the ▲ or ▼ key to select
Initialize Setting
, and press
ENTER
.
The pressure measurement settings are initialized.