3-14
IM CA700-01EN
Saving the Data after Adjustment
16.
In the calibration display, press
As Left (F5)
.
The display switches to the source and measure value display.
Current calibration point/
total number of calibration points
Error at the current calibration point
17.
Using the same procedure as in “Saving the Data before Adjustment,” save the calibration data
after adjustment.
18.
When you finish saving the data of all calibration points, press
Done (F5)
.
The calibration data of all calibration points are saved to a single CSV file, and the calibration result is
shown on the display.
19.
Press
Done (F5)
.
The calibration display returns.
3.4 Calibrating the Pressure Transmitter