Saving Calibration Data
You can save calibration data to a file. If you adjust a transmitter, the data before and after the
adjustment can be saved to files with the following file names.
File Name
The name of the file for data before adjustment is MSxxF_y.CSV where
MS: Fixed (measurement/source file)
xx: Calibration group number (00 to 99)
F: Fixed (data before adjustment)
y: File number (0 to 9)
The name of the file for data after adjustment is MSxxL_y.CSV where
MS: Fixed (measurement/source file)
xx: Calibration group number (00 to 99)
L: Fixed (data after adjustment)
y: File number (0 to 9)
The calibration group number remains the same as long as the calibration target information
or calibration procedure remain unchanged. Therefore, the calibration data files of the same
transmitter before and after adjustment will have the same calibration group number.
Saved Information
The following information is saved in a calibration data file.
• Instrument information
• Calibration procedure information
• Calibration date
• CA700 serial number
• Calibration result
Pressure Switch Calibration
In the calibration procedure in “Transmitter Calibration,” if you select pressure switch (P SWITCH, EXT
SWITCH) for measurement, you can calibrate pressure switches.
The CA700 measures the pressure required for the switch to open and close, the difference between
the open and close pressures, and the resistance when the switch is closed.
The CA700 can calibrate mechanical pressure switches. It cannot be used to calibrate open-collector-
contact, voltage-output, or current-output pressure switches.
Saving Calibration Data
Like transmitter calibration, the calibration results can be saved.
File Name
The name of the file for data before adjustment is SWxxF_y.CSV where
SW: Fixed (pressure switch test data file)
xx: Calibration group number (00 to 99)
F: Fixed (data before adjustment)
y: File number (0 to 9)
The name of the file for data after adjustment is SWxxL_y.CSV where
SW: Fixed (pressure switch test data file)
xx: Calibration group number (00 to 99)
L: Fixed (data after adjustment)
y: File number (0 to 9)