<5. Functions>
96
IM 01E21A02-03EN
NE
107
Error Message
Error Description
Countermeasure
FD bit
Display
F
OUNDATION
Fieldbus
C 361:DiagTB O/S Mode Diagnostic TB in
O/S Mode
The DTB block is in the
O/S mode.
Change the DTB block
mode.
7 14
C 362:MaintTB O/SMode Maintenance TB in
O/S Mode
The MTB block is in the
O/S mode.
Change the MTB block
mode.
7 13
C 363:DispTB O/S Mode Display TB in O/S
Mode
The LTB block is in the
O/S mode.
Change the LTB block
mode.
7 12
Warning
The device works normally and measurement is also normal but warning occurs.
NE
107
Error Message
Error Description
Countermeasure
FD bit
Display
F
OUNDATION
Fieldbus
S 082:Pls 1 saturate
Pulse output 1
saturated
Saturation of pulse
output 1 was detected.
Check process value
and pulse output 1
parameter settings.
3
5
C 085:Cable miscon
Cable misconnect
Misconnection of cable
was detected.
Check the signal cable
and excitation cable
connection.
3
2
C 086:Coil insulation
Coil insulation
warning
Insulation deterioration
of the coil was
detected.
Contact Yokogawa
service center.
3
1
M 087:Adhesion lv 3
Adhesion over level
3
The resistance value
of the electrodes
exceeded Level 3.
(Adhesion detection
of insulation to the
electrode)
Recommend cleaning
the electrode.
4 31
N 088:LC warn
Low conductivity
warning
Decrease of
conductivity was
detected.
Check fluid
conductivity.
4 30
M 089:Insu detect
Insulation detection
Insulation deterioration
of electrode was
detected.
Contact Yokogawa
service center.
4 29
N 090:FLN over lv 3
Flow noise over
level 3
Flow noise exceeded
Level 3.
(Detection of flow
noise)
Check if the fluid has a
problem (conductivity
and bubble).
4 28
N 091:FLN over lv 4
Flow noise over
level 4
Flow noise exceeded
Level 4.
(Detection of flow
noise)
Check if the fluid has a
problem (conductivity
and bubble).
4 27
C 092:AZ warn
Autozero warning
Result of zero
adjustment exceeded
10 cm/s.
Check fluid is stopped
when executing zero
adjustment.
4 26
C 093:Verif warn
Verification warning
Interruption of
verification function
was detected.
Execute the Verification
function again.
4 25
C 094:Fact noise warn
Factory noise
warning
The fluctuation of flow
became larger.
Check the fluid.
4 24
C 095:Simulate active
Simulation active
The test mode for any
of the flow velocity,
volumetric flow rate,
mass flow rate, calorie,
pulse output, status
input, or status output
was executed.
Release simulation or
test mode.
4 23
S
098:Pls 1 fix
Pulse output 1 fixed
It was detected that
pulse output 1 is fixed.
Check whether pulse
output 1 is in the test
mode or not.
4 20