<5. Functions>
54
IM 01E21A02-03EN
Flow noise
Menu path
Display
Device setup ► Diag/Service ► Diagnosis ► Flow noise ► Result ► (see below)
F
OUNDATION
Fieldbus
Process Variables ► DTB ► Process Variables ► Dynamic Variables ► Device Variables
► Flow Noise Value ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
DTB
20
-
Flow Noise Value.Status
Status of the flow noise value
Value
Flow Noise Value.Value
Displays the flow noise value.
This parameter can be used only for AXG, not for AXW.
Electrode adhesion detection
Menu path
Display
Device setup ► Diag/Service ► Diagnosis ► Adhesion ► Result ► (see below)
F
OUNDATION
Fieldbus
Process Variables ► DTB ► Process Variables ► Dynamic Variables ► Device Variables
► Adhesion Value ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
DTB
14
-
Adhesion Value.Status
Status of the resistance value
of the electrode
Value
Adhesion Value.Value
Displays the resistance value of
the electrode.
Electrode potential
Menu path
Display
Device setup ► Diag/Service ► Diagnosis ► Empty check ► (see below)
F
OUNDATION
Fieldbus
Process Variables ► DTB ► Process Variables ► Dynamic Variables ► Device Variables
► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
DTB
16
-
Electrode A Value ► Electrode
A Value.Status
Status of voltage value
between electrode A and
electrode C
Value
Electrode A Value ► Electrode
A Value.Value
Displays the voltage value
between electrode A and
electrode C.
DTB
18
-
Electrode B Value ► Electrode
B Value.Status
Status of voltage value
between electrode B and
electrode C
Value
Electrode B Value ► Electrode
B Value.Value
Displays the voltage value
between electrode B and
electrode C.