<5. Functions>
143
IM 01E21A02-03EN
5.10.4 Test Mode Auto Reset
If the specified time lapses with no parameters changed while the test mode is enabled, the test
mode is reset automatically. When any test mode parameter is changed, the test mode reset time
is extended.
The setting can be configured with the following parameter.
Menu path
Display
Device setup ► Diag/Service ► Test ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► MTB ► Device Configuration ► Query Device ► MaintTB Original
Parameters(Part2) ► Page 4 ► Test
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
MTB
95
Release time
Test Auto Release Time
Specifies the time to
automatically reset the test
mode.
From the table below, select the test mode auto reset time.
Selection
Description
Display
F
OUNDATION
Fieldbus
10 min
10min
Sets the reset time to 10 min.
30 min
30min
Sets the reset time to 30 min.
1 h
1h
Sets the reset time to 1 hour.
3 h
3h
Sets the reset time to 3 hours.
12 h
12h
Sets the reset time to 12 hours.