<5. Functions>
63
IM 01E21A02-03EN
5.1.10 Temperature Setting
The temperature setting is required when calculating the calorie based on the temperature
difference from the temperature that is taken in via MAO FB.
For details about the calorie measurement, refer to Subsection 5.5.6.
This parameter can be used only for AXG, not for AXW.
The setting can be configured with the following parameters.
Menu path
Display
Device setup ► Detailed setup ► Pro var ► Temperature ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► STB ► Device Configuration ► Configuration ► Process
Variables ► Temperature ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
STB
101
Std temperature
Standard Temperature
Specifies the reference
standard temperature for using
the temperature-based density
correction function.
STB
107
Fixed temperature
Calorific Fixed Temp
Specifies the reference
temperature to use for the
calorie calculation with
temperature difference from the
temperature which is input from
the outside.
Display
Device setup ► Detailed setup ► Pro var ► Temperature ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► STB ► Device Configuration ► Query Device ► SnsrTB Original
Parameters(Part2) ► Page 1 ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
STB
104
Meas temperature
Measured Temperature
Displays the temperature that is
input from the outside.