<5. Functions>
134
IM 01E21A02-03EN
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
DTB
46
Diagnostic output
Diagnostic Output
Specifies the output to execute
the verification function.
*1
Menu path
Display
Device setup ► Diag/Service ► Verification ►(see below)
F
OUNDATION
Fieldbus
Diagnostic ► DTB ► Device Diagnostics ► Diagnostics/Alerts ► Maintenance ►
Verification ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
DTB
48
Mode
Verification Mode
Specifies the fluid status to execute
the verification function.
*2
DTB
49
Execute
Verification Exe
Specifies the execution of the
verification function.
*3
DTB
50
VF No
Verification No
Specifies the diagnosis result
display time.
*4
DTB
47
VF target select
Verification Target
Specifies the target for diagnosis.
*5
*1:
From the table below, select the output to execute the verification function.
Selection
Process value
Totalizer
Pulse output
Frequency
output
Display
F
OUNDATION
Fieldbus
Zero
Zero
0% output
Input 0
(Output fixed)
0 pps
0% output
Measured value Measured value
Output of
calculated value
(Undefined)
Totalization of
calculated value
(Undefined)
Output of
calculated value
(Undefined)
Output of
calculated value
(Undefined)
Hold
Hold
Fixes the last
valid value
before the
diagnosis starts.
Totalizes the
last valid value
before the
diagnosis starts.
Fixes the last
valid value
before the
diagnosis starts.
Fixes the last
valid value
before the
diagnosis starts.
*2:
From the table below, select the fluid state to execute the verification function.
Selection
Description
Display
F
OUNDATION
Fieldbus
No flow
No flow
Executes the verification function with the fluid not flowing.
Flow
Flow
Executes the verification function with the fluid flowing.
*3:
From the table below, select whether or not to execute the verification function. In the communication access of
F
OUNDATION
Fieldbus, the communication access is set with the procedures of the interactive operation guide called
DD Method.
Selection
Description
Display
F
OUNDATION
Fieldbus
Not execute
Not execute
Does not execute the verification function.
Execute
Execute
Executes the verification function.
*4:
From the table below, select the execution result of the verification function.
Selection
Description
Display
F
OUNDATION
Fieldbus
Factory
Factory
Displays the result obtained at shipment from the manufacturing
factory.
Previous
Previous
Displays the previous result.
Present
Present
Displays the result at this time.