<5. Functions>
168
IM 01E21A02-03EN
5.14.6 Frequency Output Range Setting
When the frequency output is used, the frequency at 0% and 100% can be specified for the span
of the process value. The frequency output range can be set by specifying the frequency.
Set the output frequency for the span of the process value which is assigned to PV in Subsection
5.1.2.
The setting can be configured with the following parameters.
Menu path
Display
Device setup ► Detailed setup ► Pulse/Status out ► PO1/SO1 ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► MTB ► Device Configuration ► Maintenance ► Pulse/Status Out
► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
MTB
110
Frequency at 0%
Frequency 1 at 0%
Specifies the output frequency
when the process value is 0%.
MTB
111
Frequency at 100% Frequency 1 at 100%
Specifies the output frequency
when the process value is
100%.
5.14.7 Status Output Function Setting
When the status output is used, the device status can be output as a contact point.
The setting can be configured with the following parameter.
Menu path
Display
Device setup ► Detailed setup ► Pulse/Status out ► PO1/SO1 ► (see below)
F
OUNDATION
Fieldbus
Device Configuration ► MTB ► Device Configuration ► Maintenance ► Pulse/Status Out
► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
MTB
112
SO1 function
SO1 Function
Specifies the status output
function.
From the table below, select the status output function.
Selection
Description
Display
F
OUNDATION
Fieldbus
No function
No function
The status output is not available because the status output function is
not enabled.
Alarm output
Alarm output
The status output is activated when the specific alarm occurs.
Warning output
Warning output
The status output is activated when the specific warning occurs.
Total limit 1
Total limit 1
The status output is activated while the totalized value of totalizer 1
is out of the set target value. For details about the totalization switch
function, refer to Subsection 5.2.3.
Total limit 2
Total limit 2
The status output is activated while the totalized value of totalizer 2
is out of the set target value. For details about the totalization switch
function, refer to Subsection 5.2.3.