<5. Functions>
135
IM 01E21A02-03EN
*5:
Select the target for diagnosis from the table below (All of them are set to diagnosis target at the time of shipment).
Selection
Description
Display
F
OUNDATION
Fieldbus
Magnetic circuit
Magnetic
Magnetic circuit diagnosis
Excitation circuit
Excitation
Excitation circuit diagnosis
Calculation circuit
Calculation
Calculation circuit diagnosis
Device status
Device status
Device alarm diagnosis
Connection status
Conn status
Wiring misconnection check
In the default state, all of the diagnosis results are reflected in the final results of the verification.
Even if set with “VF target select” so as not to reflect the diagnosis results to the final results of
the verification, execution time for the verification does not change because the diagnosis itself is
executed. In addition, the diagnosis results selected to “0” with Verification Target are displayed
as “Skip”.
Note that verification itself cannot be executed if all diagnoses are unchecked in the check box of
Verification Target (0x0000).
The diagnosis result is displayed as shown below.
Display
Device setup ► Diag/Service ► Verification ► Result ► (see below)
F
OUNDATION
Fieldbus
Diagnostic ► DTB ► Device Diagnostics ► Diagnostics/Alerts ► Maintenance ►
Verification ► (see below)
Block
Name
Relative
Index
Parameter
Description
Display
F
OUNDATION
Fieldbus
DTB
51
Failed/Passed
VF Check Results
Execution result
DTB
52
VF operate time
VF Operation Time
Operation time of
verification when
started
DTB
53
Magnetic circuit
Magnetic Circuit Res
Magnetic circuit
diagnosis result
DTB
54
Excite circuit
Exciting Circuit Res
Excitation circuit
diagnosis result
DTB
55
Calc circuit
Calc Circuit Res
Calculation circuit
diagnosis result
DTB
56
Device status
Device Status Res
Device alarm
diagnosis result
DTB
57
Connect status
Connection Status Res
Wiring connection
diagnosis result
The diagnosis result is judged as shown below.
Result
Description
Display
F
OUNDATION
Fieldbus
Passed
Passed
There are no problems concerning the diagnosis result.
Failed
Failed
There is a problem concerning the diagnosis result.
Canceled
Canceled
Cancels the diagnosis.
No data
No Data
No diagnosis result data (The verification function is not used.)
Unknown
Unknown
Cannot perform a diagnosis.
Skip
Skip
Out of verification target
IMPORTANT
• Before using the verification function, be sure to disconnect this product from the control loop.
• While the verification function is used, the pulse output is not output correctly.
• Note that parameters cannot be changed while the verification function is executed.