8. Setting of parameters
105204
– WTS-300-E - V4.1 - 20.01.2022
Page 99
Section 8.6
: Level “Extended cleaning parameters 2” (L parameters)
Settings for quantity-controlled water intake
L104 Water quantity in dosing position 1
only active if quantity-controlled water intake is
activated
0…50 l
5
L105
Activation of quantity-controlled water
intake
0: deactivated
1: activated
0…1
0
L106 Timeout settings for water intake
Same timeout (L107) for all programme steps
0: After timeout stop of the cleaning with fault
message
1: Continue cleaning after timeout with fault
message
Individual timeout for each programme step (L107
plus step time from the corresponding n-parame-
ter)
2: Continue cleaning after timeout and report fault
0…2
1
L107 Timeout time for water intake
1,0..25,0
min.
5,0
Settings for quantity-controlled detergent and disinfectant
dosing
L110 Activation of dosing of quantity-controlled
detergent and disinfectant
0: deactivated
1: activated
0…1
0
L112
Dosing of timeout-time
0,1..10,0
min.
1,0
Setting of the pulse counters: Pulses per litre
L121
Impulse counter “acidic“
1…9999
2000
L122
Impulse counter “alkaline“
1…9999
2000
L123
Impulse counter “disinfectant“
1…9999
2000
L124
Impulse counter “cold water“
1…9999
2000
L125
Impulse counter “warm water“
1…9999
2000