23
Explanation of command
4.2.33
DTIMER=
(setting of test time of DC withstanding voltage test)
Function
Makes the setting of test time of DC withstanding voltage test.
Structure
DTIMER=
Test time
Test time OFF or 0.5~999sec. is to be set.
Transmission
DTIMER=60.0s
…….. Sets the test time of DC withstanding voltage test at
60.0sec.
Response
When 8526 received the effective command setting.
ERROR=0
…….……. When the Response Setting is ON.
No response …………….. When the Response Setting is OFF.
4.2.34
DTIMER?
(read-out of test time of DC withstanding voltage test)
Function
Reads out the test time of DC withstanding voltage test.
Structure
DTIMER?
Transmission
DTIMER?
Response
DTIMER=10.0s
…….. Indicates the test time of DC withstanding voltage test
10.0sec.