Setting of test condition for withstanding voltage test
[When setting the low limit of leak current]
To enter setting mode
①
In READY status, press
or
key, then the test mode lamp blinks. The test
mode lamp moves up and down with
▲
or
▼
key. Make the required test
mode lamp blinking (
AC
or
DC
).
②
Press
or
key and make the test mode lamp lit up.
Setting of low limit of leak current
①
Press
or
key and select the status that the current display blinks and
LOW SET
is lit up.
②
Next, press
ON/OFF
key and select the status that the display blinks with the
numeral
(refer to the above figure).
③
Press
▲
or
▼
key and set the low limit of leak current value.
A press of
key (
SHIFT
and
▲
keys at a time) or
key (
SHIFT
and
▼
keys at a time) allows the setting of second digit.
To move to the previous setting
Press
key, then changes to the
setting of high limit of leak current
.
To the next setting
Press
key, then changes to the
setting of test time
.
Finish of setting
Press
ENTER
key, then the tester returns to READY status, memorizing the settings
having been made.
When the
EXIT
key (
SHIFT
and
ENTER
key at a time) is pressed in the setting
mode, the setting mode for the test condition is interrupted and the tester becomes
READY
status.
The test conditions in this case are those before entering the setting mode of test
condition.