2
2. Confirmation prior to use
2.1
●
Unpacking
(1)
Unpacking
When the tester is delivered, please check whether it has not been damaged in transit
and unpack it carefully. If any damage or inconvenience is found, please consult the
dealer whom you purchased the tester from for proper solution.
(2) Check of contents
Please do not leave in the carton any item of the contents listed below.
List of accessories:
High voltage cable 2m
1 pair
Earth wire 3m
1 piece
Power supply cord 2.5m
1 piece
REMOTE/OUT plug
1 piece (36P)
Fuse 7A
1 piece
Instruction manual
1 copy
RS-232C interface instruction manual
1 copy
RS-232C connector (D-sub 9 pins) Model 5881-11-020 (9 pins – 9 pins / 2.0m) for
external communication is available at option. When a customer procures it,
please use the inch pitch screw type.
2.2
●
Cautions for handling
Since the Model
8526
deals high voltage, it is designed paying special attention to safety.
However, it is still dangerous as it outputs high voltage of max. 5kV. An erroneous
handling may cause fatal accident. In order avoid any accident, please strictly observe
the following cautions and take utmost care for safety.
(1) Make sure to connect the protective grounding terminals (back panel) to the earth. If
the grounding is insufficient, the tester housing is charged with high voltage when the
output is short-circuited to the earth or the power source line, and is very dangerous.
Please also check if the grounding cable is disconnected or not.
WARNING
Insufficient grounding may cause the electric shock.
(2) Never touch the output terminals, high voltage cable and test samples during the test.
(3) When making a connection to the test sample, connect the LOW side prior to the other,
with the output OFF.
(4) When operating the Model
8526
, put a rubber glove for prevention of electric shock.
CAUTION