14. External control
14.1
●
Control by REMOTE/OUT connector
By means of the
REMOTE/OUT
connector
⑱
on the rear panel, the remote control of
start/stop, the setting of interlock to secure the safety, and the output signals corresponding
to each condition of the
8526
can be output by open collector.
The input and output signals are isolated from the internal circuit by photo-coupler.
Also, the
8526
is provided with the power source of 24V DC 0.1A, which can be utilized
as power supply for the external control.
14.2
●
Arrangement and function of connector pins
I/O
Signal name
Pin No.
Function
+24V
1
Power 24V DC for external control is output.
(capacity 0.1A)
I
REAR:ON
2
Change-over signal for remote control.
Ref. article 13.3 for detail.
START
3
Input signal for start.
STOP
4
Input signal for stop.
INTERLOCK
5
Signal for interlock.
MEM SET 1
6
BCD code input for read out of memory.
(effective at the setting of REAR:MEM)
Effective for No.1~No.9
A~F code are ineffective, no memory can be read.
MEM SET 2
7
MEM SET 4
8
MEM SET 8
9
O
TEST/H.V.OUT
10
Output at high voltage terminal during the voltage
output.
READY
11
Output at READY status.
PROTECTION
12
Output when the protective function works.
Ref. article 14.4 for detail.
GOOD
13
Output at good judgement.
HIGH
14
Output at NG judgement for high limit.
-
NC
15
Vacant pin (do not use it as relay terminal).
NC
16
Vacant pin (do not use it as relay terminal).
NC
17
Vacant pin (do not use it as relay terminal).
NC
18
Vacant pin (do not use it as relay terminal).
COM COM
19
Common (common with 23, 36)
I
REAR:MEM
20
Change-over signal for memory read out from the
rear panel.
-
NC
21
Vacant pin (do not use it as relay terminal).
NC
22
Vacant pin (do not use it as relay terminal).
COM COM
23
Common (common with 19, 36)
O
AC-TEST
24
Output during the AC withstanding voltage test.
Not output while
AC
is blinking.
DC-TEST
25
Output during the DC withstanding voltage test.
Not output while
DC
is blinking.
TEST
26
Output during the test. Not output while
AC
or
DC
is blinking.
-
NC
27
Vacant pin (do not use it as relay terminal).
O
END
28
Output at the end of test.
-
NC
29
Vacant pin (do not use it as relay terminal).
NC
30
Vacant pin (do not use it as relay terminal).
O
NG
31
Output at NG judgement.
LOW
32
Output at NG judgement for low limit.
-
NC
33
Vacant pin (do not use it as relay terminal).
NC
34
Vacant pin (do not use it as relay terminal).
NC
35
Vacant pin (do not use it as relay terminal).
COM COM
36
Common (common with 19, 23)
Type of input/output:
I: input
O: Open collector output.
COM: Common for input/output
-
: Vacant pin