![Tsuruga 8526 Instruction Manual Download Page 80](http://html1.mh-extra.com/html/tsuruga/8526/8526_instruction-manual_1178554080.webp)
28
Explanation of command
4.2.41
MEM No :
(setting of test condition to memory)
Function
Makes the setting of test mode and parameters in the designated
memory number.
Structure
MEM No :
Parameter of test
No
: 1~9
Parameter of test
Same as those at the
article 4.2.37 (P26) SET: (setting of parameters
of test condition)
Transmission
In the AC withstanding voltage test mode.
MEM5:MODE=AC,AVOLT=5.0kV,ALEVEL=1.00kV,AHIGH=100.0mA,ALOW=OFF,
ATIMER=60.0s
In the DC withstanding voltage test mode.
MEM5:MODE=DC,DVOLT=5.0kV,DLEVEL=1.00kV,DHIGH=10.0mA,DLOW=OFF,
DTIMER=60.0s
Response
When 8526 received the effective command setting.
ERROR=0
…….……. When the Response Setting is ON.
No response …………….. When the Response Setting is OFF.
4.2.42
MEM No :?
(read-out memorized test condition)
Function
Reads out the designated memory number, test mode and each
parameter in the lump.
Structure
MEM No :?
Transmission
No
: 1~9
MEM3:?
Response
In the AC withstanding voltage test mode.
○
When
FORMAT=ON
MEM3:MODE=AC,AVOLT=2.5kV,ALEVEL=1.50kV,AHIGH=20.0mA,ALOW=OFF,
ATIMER=60.0s
○
When
FORMAT=OFF
MEM3:AC,2.5,1.50,20.0,OFF,60.0
In the DC withstanding voltage test mode.
○
When
FORMAT=ON
MEM3:MODE=DC,DVOLT=2.5kV,DLEVEL=1.50kV,DHIGH=10.0mA,DLOW=OFF,
DTIMER=60.0s
○
When
FORMAT=OFF
MEM3:DC,2.5,1.50,10.0,OFF,60.0