22
Explanation of command
4.2.31
DLOW=
(setting of low limit of leak current of DC withstanding voltage test)
Function
Makes the setting of low limit of leak current of DC withstanding
voltage test.
Structure
DLOW=
Low leak current
Low leak current OFF or 0.0~10.9mA is to be set.
Note
: Set value of low leak current can not be higher than high limit
value of leak current.
Transmission
DLOW=2.0mA
…….. Sets the low limit of leak current of DC withstanding
voltage test at 2.0mA.
Response
When 8526 received the effective command setting.
ERROR=0
…….……. When the Response Setting is ON.
No response …………….. When the Response Setting is OFF.
4.2.32
DLOW?
(read-out of low limit value of leak current of DC withstanding voltage test)
Function
Reads out the low limit value of leak current of DC withstanding
voltage test.
Structure
DLOW?
Transmission
DLOW?
Response
DLOW=2.0mA
…….. Indicates the low limit of leak current of DC withstanding
voltage test 2.0mA.