![Tsuruga 8526 Instruction Manual Download Page 38](http://html1.mh-extra.com/html/tsuruga/8526/8526_instruction-manual_1178554038.webp)
33
External control
14.5
●
Output signals and power supply for control
It is possible to take out each condition of the
8526
as output signal.
The power supply of 24V DC for control is provided, so the relay etc. can be directly
driven.
(1) Specification of output signal (Pin No.10~14, 24~26, 28, 31, 32)
Signal type
: Open collector output
Max. load voltage
: 30V DC
Max. output current
: 30mA DC
Isolation system
: Isolated from the internal circuit by photo-coupler
Output saturation voltage : 1.6V DC or less
(2) Specification of control power source (Pin No.1)
Output voltage
: 24V DC
Current capacity
: 0.1A DC
Fig.14.2
Connection example of relay drive
Fig.14.3
Example to obtain a signal level
CAUTION
●
Use the output signal with 30V and 30mA DC or less.
●
In case of controlling an inductive load like relay, connect a diode in
parallel with the coil to absorb the reverse electricity.
PIN No.
19
、
23
、
36
PIN No.
28
、
31
、
32
10
~
14
、
24
~
26
、
Relay
COM
+
24V
Each output signal
1
8526
Each output signal
PIN No.
19
、
23
、
36
PIN No.
28
、
31
、
32
10
~
14
、
24
~
26
、
COM
+
24V
1
8526