21
9. Memory function
This tester is provided with 9 program memories to memorize the test mode and the setting of
test conditions.
9.1
●
Configuration of memory
Each memory can memorize one type of test mode, 5 items of the test conditions of AC
withstanding voltage test and 5 items of the test conditions of DC withstanding voltage test.
For the content of each type and item, please refer to the following table.
Type of test mode
Setting of test condition for
AC withstanding voltage test
Setting of test condition for
DC withstanding voltage test
↓
↓
↓
1 type
5 items
5 items
AC (AC withstanding
voltage test)
DC (DC withstanding
voltage test)
Either one
Range of test voltage
Range of test voltage
Referential voltage
Referential voltage
High limit of leak current
High limit of leak current
Low limit of leak current
Low limit of leak current
Test time
Test time
9.2
●
Memory write-in
Selection of memory No.
①
Make the setting of test mode and test condition required to be written in the memory
and make the tester
READY
status (refer to the article 7~8).
②
Press
WRITE
key (
SHIFT
and
at a time).
Then the numeral on the memory No. display blinks, entering into the memory
write-in mode (refer to the above figure).
③
Select the memory No. to write in with
▲
or
▼
key.
Finish of memory write-in
Press
ENTER
key, then the tester returns to
READY
status, memorizing the settings
having been made.
When the
EXIT
key ((
SHIFT
and
ENTER
key at a time) is pressed in the setting
mode, the memory write-in mode is interrupted and the tester becomes
READY
status.
The memory No. in this case is that before entering the memory write-in mode.
ON/OFF
ENTER
WRITE
READ
LOCK
EXIT
SHIFT
BUZZER
PROTECT
○
ONLINE
○
○
○
○
○
KEY LOCK
DISCHARGE
REMOTE
READY
AC
DC
○
No.
s
mA
SET
SET
LOW
HIGH
GOOD
LOW
MODE
JUDGE
CURRENT
TEST VOLTAGE
MEMORY
RANGE
TIME
kV
kV
DC
AC
HIGH
●