29
Remote control
Specification of input signal
:
Control input: Active LOW
Input level: “H”=16.8~24V
“L”=0~3.8V
“L” level flow out current: I
C
=10mA
“L” level min. pulse width: 40ms
Note
: START, STOP terminals are
pulled up to +24V, so they
become “H” level at opening.
Fig.13.1
Connection examples of remote control terminal
CAUTION
In case that the control is made by switch, relay and etc. and when the
chattering occurs, it may cause faulty operation.
13.3
●
Operation by REMOTE/OUT connector
Same remote operation as that through
REMOTE
terminal
can be done through the
REMOTE/OUT
connector
⑱
on the rear panel.
For connection of connector, please refer to the article 14.2 (P31).
The operation is same as that of
REMOTE
terminal, the article 13.2 (P28).