14
Explanation of command
4.2.14
IDNT?
(read-out of tester identification)
Function
Reads out the model name, software version of the tester.
Structure
IDNT?
Transmission
IDNT?
Response
When 8526 received the effective command setting.
IDNT=TSURUGA_8526_ROM-NO.456_Ver.1.00.00
① ②
①
Model name
②
Software version
(For improvement of quality, the software version might have been updated.)
4.2.15
AVOLT=
(setting of test voltage range of AC withstanding voltage test)
Function
Makes the setting of test voltage range of AC withstanding voltage test.
Structure
AVOLT=
Test voltage range
Test voltage range 2.5kV or 5.0kV is to be set
Transmission
AVOLT=5.0kV
…….. Sets the range of AC withstanding voltage test at 5.0kV.
Response
When 8526 received the effective command setting.
ERROR=0
…….……. When the Response Setting is ON.
No response …………….. When the Response Setting is OFF.
4.2.16
AVOLT?
(read-out of test voltage range of AC withstanding voltage test)
Function
Reads out the test voltage range of AC withstanding voltage test.
Structure
AVOLT?
Transmission
AVOLT?
Response
AVOLT=2.5kV
….…. Indicates the test voltage range of AC withstanding voltage
test 2.5kV.