10
5. Setting items in each mode
5.1
●
READY status
When the
POWER
switch
①
is turned ON, the lamp test is done and then
READY
is
lit up showing that the tester enters in
READY
status.
The test condition at the previous power shutdown is displayed.
Pressing the
START
switch
③
starts the test.
In READY mode, the setting of the following items can be done.
Items to set
(1) Test condition
Refer to the article 7~8 (P12~20)
(2) Key lock
Refer to the article 11 (P26)
(3) Buzzer sounding
Refer to the article 17 (P37)
(4) Status output condition
Refer to the article 15.3 (P35)
(5) Special test mode
Refer to the article 12 (P27)
①
Double action
②
GOOD hold
③
Momentary
④
FAIL mode
5.2
●
Setting mode of test condition
In READY status, by pressing the
(or
) key,
READY
is turned off and the
tester enters into the test condition setting mode.
In the test condition setting mode, the test mode and condition can be set or changed.
A press of
ENTER
key finishes the setting and the tester becomes
READY
status.
Item to set
Test mode
Test condition
Ref. art. 7.2 (P12)
Ref. art. 8 (P13~20)
5.3
●
Memory write-in mode
After setting the test condition in the test condition setting mode, press the
WRITE
key
(
SHIFT
+
), then the memory number blinks, being ready to write in the memory.
In the memory write-in mode, 9 memory sets can be written. Each memory set consists
of 6 items of test conditions which are set in the test condition setting mode.
A press of
ENTER
key finishes the setting and the tester becomes
READY
status.
Item to set
Test condition
→
Memory write-in (No.1)
→
Test condition
→
Ref. art. 7.2 (P12) Ref. art. 9.2 (P21) Ref. art. 7.2 (P12)
Memory write-in (No.2)
Test condition
→
Memory write-in (No.9)
Ref. art. 9.2 (P21) Ref. art. 7.2 (P12) Ref. art. 9.2 (P21)
5.4
●
Memory read-out mode
In READY status, by pressing the
READ
key
(
SHIFT
+
), a memory No. blinks
and the tester becomes ready to read out the memory. In the memory read out mode, one
of the 9 memories written in [ref. art. 9.2 (P21)] can be called up and read out.
A press of
ENTER
key finishes the setting and the tester becomes
READY
status.
Item to set
Memory read out
(Select memory No.)