27
12. Special test mode
Model
8526
is able to have the setting of 4 special functions by means of key operation on
the front panel.
(1) Double action start function
Within 0.5 second from the stop signal having been ON/OFF, the test starts with input
of start signal.
Note
: When the function is set,
READY
lamp blinks in READY status.
(2) GOOD hold function
This is the function to concern the good judgement. The output becomes continuous
until the stop signal is input.
(3) Momentary start function
The test is done only when the start signal is input.
(4) FAIL mode function
This is the function to disable the resetting of NG judgement and PROTECTION
action by the stop signal of remote control, and enables the resetting only by the stop
switch on the tester main unit.
Setting procedure of special test mode
①
In READY status, press
SHIFT
key and
STOP
key at a time for 3 seconds or
more.
READY
lamp blinks and the test time display is lit up with “
”.
The highest digit of the current display blinks.
②
The item to set can be moved with
or
key.
③
Refer to the following table for the items to select.
CURRENT
▲
key: Numeral increases.
▼
key: Numeral decreases.
Lamps to synchronously blinks
at the setting
-
-
-
Cancel of setting
READY
lamp
-
-
-
Setting of double action start function
-
-
Cancel of setting
GOOD
-
-
Setting of GOOD hold function
Note
: In order to re-start, once of stop
signal input is necessary
-
-
Setting of GOOD hold function
Note
: When the start signal is input, the
judgement output is reset and
re-starts.
-
Cancel of setting
AC
DC
-
Setting of momentary start function
Cancel of setting
HIGH
LOW
Setting of FAIL mode
Finish of setting
Press
ENTER
key, then the setting is memorized and returns to READY status.
When the
EXIT
key (
SHIFT
and
ENTER
key at a time) is pressed in the setting mode,
the special test mode is interrupted and the tester becomes
READY
status.
The special test mode in that case is the condition before entering the special test mode.
ON/OFF
ENTER
WRITE
READ
LOCK
BUZZER
EXIT
SHIFT
PROTECT
○
ONLINE
○
○
○
○
○
KEY LOCK
DISCHARGE
REMOTE
READY
AC
DC
○
○
No.
s
mA
HIGH SET
SET
LOW
HIGH
GOOD
LOW
MODE
JUDGE
CURRENT
TEST VOLTAGE
MEMORY
RANGE
TIME
kV
kV
AC
DC
Double action
GOOD hold
Momentary
FAIL mode