18
Explanation of command
4.2.23
ATIMER=
(setting of test time of AC withstanding voltage test)
Function
Makes the setting of test time of AC withstanding voltage test.
Structure
ATIMER=
Test time
Test time OFF or 0.5~999sec. is to be set.
Transmission
ATIMER=60.0s
…….. Sets the test time of AC withstanding voltage test at
60.0sec.
Response
When 8526 received the effective command setting.
ERROR=0
…….……. When the Response Setting is ON.
No response …………….. When the Response Setting is OFF.
4.2.24
ATIMER?
(read-out of test time of AC withstanding voltage test)
Function
Reads out the test time of AC withstanding voltage test.
Structure
ATIMER?
Transmission
ATIMER?
Response
ATIMER=10.0s
…….. Indicates the test time of AC withstanding voltage test
10.0sec.