25
Test procedure
※
Caution When the test voltage is out of the range of referential voltage
[When the test voltage is out of the range of referential voltage]
①
In case that the referential voltage is set and when the test voltage is not within the
range of referential voltage (within 5% of the set value), the test is stopped. (For the
voltage less than 1000V, /-50V (+/-5 digit).)
When the test voltage is less than the range of referential voltage, the tester waits for 5
seconds, and when exceeded, the tester immediately stops the test.
Also, when the test voltage went out of the range of referential voltage, the test is
immediately stopped.
②
Press
START
switch
③
, then the
DANGER
lamp
⑩
is lit up, and if the test
voltage is our of the range of referential voltage,
MODE AC
or
DC
blinks.
③
During this sequence, the test voltage display displays the measured output voltage
value and the current display displays the measured leak current value. Also, if the test
time is set, the test time display displays the set value, and when the test time is set to
OFF, the test time display displays
s.
④
At the
REMOTE/OUT
connector
⑱
, TEST/H.V.OUT is ON.
⑤
When judged to be out of the referential voltage range and the test is stopped, the
current display displays the value at that moment, and the test voltage display displays
the value at that moment in blinking. Also,
JUDGE
HIGH
LOW
and
PROTECT
lamp are lit up.
⑥
At this time,
PROTECTION
is ON at the
REMOTE/OUT
connector
⑱
.
⑦
Press
STOP
switch
②
, then the tester resets and returns to
READY
status.