Technical Data
98
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
Adjustment ranges
Lift/lower
21.7" (55 cm) (tolerance ±0.4" (±1 cm))
Longitudinal travel
18" (46 cm) (tolerance ±0.4" (±1 cm))
Head end
7" (18 cm)
(tolerance ±0.4" (±1 cm))
Foot end
11" (28 cm)
(tolerance ±0.4" (±1 cm))
Speed
3 cm/s
Trendelenburg/reverse
Trendelenburg (adjustment
about the transverse axis)
45°/45°
Tilt (adjustment about the
longitudinal axis)
30° to the left/30° to the right
Leg section joints
Angle specifications in
relation to the seat plate
90° upwards/105° downwards
Back section joints
Angle specifications in
relation to the seat plate
Two-part table top
90° upwards/45° downwards
One-part table top
90° upwards/90° downwards
Classification
Operating table protection classification
I
Device with internal power supply I. P. S
Level of protection against electrical shock
for the entire operating table
Type B applied part
Patient leakage current in accordance with CF
in compliance with IEC60601-1
Degree of protection from water
penetration
IP X4
Operating mode
Continuous operation with intermittent load
(DBAB) 2 min ON, 8 min OFF