Safety
Instruction manual Operating table TruSystem 7000 – 1 730 263 – 11/2016
13
1.2.2
Electricity
WARNING!
Risk of personal injury.
Position the operating table such that the detachable
plug or device couplers are easily accessible.
The operating table may only be connected to a
power supply with protective conductor. This reduces
the risk of electric shock.
WARNING!
Risk of personal injury.
Have the operating table and power supply electrical
safety checked once a year by an electrician.
We recommend an annual general safety inspection
by TRUMPF Medical Systems, Inc., Technical
Customer Service.
WARNING!
Risk of personal injury. Hazard to the patient!
• Always keep the operating table in an operational (charged)
state so that it is ready immediately in case of power supply
failure!
• Fully charge the operating table prior to first use.
WARNING!
Risk of personal injury.
An extension cord cannot be used with the power
cable at the operating table, otherwise, in case of a
fault, the permitted patient leakage current threshold for
cardiac surgery (applied part CF according to
IEC 60601–1) will be exceeded. The fault occurs
when the protective conductor is interrupted by a
damaged power cable. The operating table must only
be connected to the power grid using the original
power cord from Trumpf Medical. In addition, the
supply voltage must be respected. The operating table
must only be connected to the supply voltage indicated
on the nameplate.